Photomicrography with SLR camera
The Axio Observer microscopes 5, 5 materials and 7, 7 materials permit the beam path to be switched to
100% vis (visual observation through eyepieces and an open position for activation of the right or left
sideport), 100% frontport (port pointing to the front for photomicrography) or 100% baseport (port
pointing downwards).
Since this is a 100% switching position,
simultaneous visual observation during
photography is not possible.
The V200 T2 2.5x adapter for SLR and the camera-
specific T2 adapter permit commercially available
35 mm SLR (DSLR) cameras to be connected to the
microscope.
•
Remove dust cover or camera lens from the
camera housing (Fig. 162/
1
).
•
Attach appropriate T2 adapter (Fig. 162/
2
) to
the camera housing.
•
Remove the dust cover from the V200 T2 2.5x
adapter for SLR (Fig. 162/
3
).
•
Screw the V200 T2 2.5x for SLR adapter into
the thread of the T2 adapter.
•
Remove dust cover from the camera port.
•
Attach the pre-mounted camera system to the
frontport (Fig. 162/
5
), align it horizontally and
tighten clamping screw (Fig. 162/
6
) using the 3
mm ball-headed screwdriver.
•
Select the object area to be photographed via the binocular tube.
•
After switching the beam path (Fig. 162/
7
) from visual observation to the frontport, 100 % of the
light is now available for the camera.
•
To operate the camera system, please observe the manufacturer's manual.
The photographic magnification on the film or image sensor is the product of objective
magnification, Optovar magnification and the 2.5 factor of the V200 T2 2.5x SLR camera
adapter.