4.9
Fitting the microscope stages
If you work on microscope stages with a recessed mounting frame or insert frame the microscope stages
should be mounted on the stand using spacer disks (see table).
Use the following spacer disks:
Stage
Order number:
Spacer disks
Mechanical stage 130x85 R/L for mounting
frame K
432016-9902-000
4 mm
Mechanical stage 130x85 R/L with short
coaxial drive, for mounting frame K
432047-9902-000
4 mm
Scanning stage 130x100 PIEZO
432027-9001-000
4 mm
Scanning stage XY DC 110x90 with stage
attachment Z-Piezo/ Rot.En. Rev.4
000000-0538-386
4 mm
Scanning stage 130x100 STEP
432029-9904-000
4 mm
Scanning stage 130x85 mot. P, CAN *
432031-9902-000
4 mm
Scanning stage 130x85 MAT, CAN
432034-9000-000
6 mm
Scanning stage 130x85 MAT, CAN
432034-9001-000
4 mm
* When using the scanning stage 130x85 mot. P, CAN
in combination with HD objectives, the recessed
mounting/inserting frames must be used.