Measure
This tab is only accessible if a
motorized (CAN bus) stage is used.
Otherwise, the
Start
button and a
display for the Z-distance
∆
Z are
displayed on the
Position
tab.
The
Measure
tab can be used to perform simple
distance measurements in millimeters (mm). Three
options are available for these measurements:
−
Distance between two manually set positions
−
Distance between a manually set position and a
defined position
−
Distance between two defined positions
If the Z distance shall be measured, activate the
incl. Z
button.
•
Move the stage to the initial position.
•
Press the
Start
button. On doing so, the display fields
∆
X,
∆
Y and
∆
Z will be set to zero.
All stage movements are displayed in the
∆
X,
∆
Y and
∆
Z fields.
The function of the position buttons is described under
(1) Position
above.
5.11.7.3
Definite Focus
The
Definite Focus
tab can be used to switch on
or off the focus stabilization of the Definite Focus
function using the
On
and
Off
buttons in the
Stabilization
field.
The
Current position
field displays the current
coordinates.
Fig. 114
Microscope -> XYZ -> Measure page
Fig. 115
Microscope -> XYZ -> Definite Focus
page