ACCURA Installation site requirements
34
ACCURA 20/30/15 (bellows cover) installation dimensions
4566
1651
620
3210
2700
3740
2460
2367
3067
≥
200
1564
640
4520
÷
All dimensions in millimeters, scale 1:50
600
680
≥
800
≥
800
CMM width
Length of granite plate, approx. 130 mm extra for
service unit.
Length of the CMM with covers
CMM height
Height of secured CMM assembly with shipping pal-
let.
Height of secured CMM assembly from lower edge of
granite plate.
Height between the granite plate and lower edge of
the stylus mounting cube on the VAST XT gold
probe.
With other VAST gold or RDS probe types, the height
is somewhat less.
Network connection for data backup and Internet
(Teleservice)
Power supply
Compressed air supply
TCC (standard) controller cabinet, MCC 800 (option)
➤
see Controller cabinet installation [
CMM <-> controller connection cable (cable length
approx. 3 m)
Compressed air connection on the CMM
Scanner protective field option
Weight:
ACCURA (size 20/30/15)
15140 kg
Workpiece, max.
5000 kg