ACCURA Installation site requirements
16
Required connections
–
A permanent connection must be provided for the system.
–
Alternatively, a 3-pin CEE plug (in accordance with IEC/EN 60309 for 32A) can also be used.
–
Shock-proof outlet for maintenance work.
NOTE
For maintenance work on the CMM with a permanent connection, an additional protective ground
connection and an outlet must be provided due to ESDS and VDE regulations. Max. distance to con-
troller: <1 m. (see connection diagram drawing)
Protection
Protect the connection with a circuit breaker. Alternatively, equivalent protective equipment can also be used. We rec-
ommend installing a lightning arrester, e.g. Phoenix Flashtrab FLT 25-400, for general protection against overvoltage.
Wiring plan for TCC controller cabinet (standard)
We recommend installing the connection in accordance with the following diagram.
Note:
Comply with applicable national regulations.
C16A
L
N
PE
PE
L
N
PE
Q7101
C16A / 15A
TCC
A
SC
A
Customer installation of local voltage (suggested)
Circuit breaker [C 16 A at 230-240 V], [15 A at 100-125 V]
Recommended: residual current circuit breaker
Cable (included with delivery)
Controller main switch
Shock-proof outlet
Note:
For maintenance work on the CMM