ACCURA Installation site requirements
47
Limit curves of permissible floor vibrations at the installation site
Permissible foundation acceleration for ACCURA X1600 with active damping
Note:
Acceleration values above the corresponding curve require an additional foundation.
0,1
1
10
100
1000
0
10
20
30
40
50
Vibration acceleration [mm/ s ²]
Frequency [Hz]
X axis horizontal
Y axis horizontal
Z axis vertical
Permissible foundation acceleration for ACCURA X2000 with active damping
Note:
Acceleration values above the corresponding curve require an additional foundation.
0,1
1
10
100
1000
0
10
20
30
40
50
Vibration acceleration [mm/ s ²]
Frequency [Hz]
X axis horizontal
Y axis horizontal
Z axis vertical