ACCURA Installation site requirements
39
Tolerances for the air dampers and leveling units
Maximum difference in level between the support surfaces: ±1 mm
Tolerance for center-to-center distance of the air dampers: ±10 mm
Note:
The tolerances must not be exceeded.
Floor load
Measuring range X/Y/Z [mm]
Value [N/mm
2
]
Mean surface pressure on the 4 support surfaces: 1600/2400/1500
2.5
1600/3000/1500
2.5
1600/4200/1500
3.0
Determined from CMM weight and maximum permissible workpiece weight distributed over the 4 support surfaces.