ACCURA Installation site requirements
21
ACCURA 16/30/15 installation dimensions
3740
4566
3210
2700
2627
620
2020
1927
1651
≥
200
1564
640
÷
All dimensions in millimeters, scale 1:50
600
680
≥
800
≥
800
CMM width
Length of granite plate, approx. 130 mm extra for service
unit.
CMM height
Height of secured CMM assembly with shipping pallet.
Height of secured CMM assembly from lower edge of granite
plate.
Height between the granite plate and lower edge of the sty-
lus mounting cube on the VAST XT gold probe.
With other VAST gold or RDS probe types, the height is
somewhat less.
Network connection for data backup and Internet (Teleser-
vice)
Power supply
Compressed air supply
TCC (standard) controller cabinet, MCC 800 (option)
➤
Controller cabinet installation [
CMM <-> controller connection cable (cable length approx. 3
m)
Compressed air connection on the CMM
Scanner protective field option
Weight:
ACCURA (model 16/30/15)
13200 kg
Workpiece, max.
4000 kg