ACCURA Installation site requirements
28
ACCURA 20/30/10 (bellows cover) installation dimensions
3740
1064
1151
670
2367
3067
640
2460
≥
200
4520
3620
2250
2760
÷
All dimensions in millimeters, scale 1:50
600
680
≥
800
≥
800
CMM width
Length of granite plate, approx. 130 mm extra for
service unit.
Length of the CMM with covers
CMM height
Height of secured CMM assembly with shipping pal-
let.
Height of secured CMM assembly from lower edge of
granite plate.
Height between the granite plate and lower edge of
the stylus mounting cube on the VAST XT gold
probe.
With other VAST gold or RDS probe types, the height
is somewhat less.
Network connection for data backup and Internet
(Teleservice)
Power supply
Compressed air supply
TCC (standard) controller cabinet, MCC 800 (option)
➤
see Controller cabinet installation [
CMM <-> controller connection cable (cable length
approx. 3 m)
Compressed air connection on the CMM
Scanner protective field option
Weight:
ACCURA (size 20/30/10)
15440 kg
Workpiece, max.
5000 kg