ACCURA Installation site requirements
18
Installation Dimensions/ Sample Installation/ Weights
Controller cabinet installation
NOTE
Pay attention to the following when installing the controller cabinet:
The controller cabinet must be positioned so that the exhaust air from the controller cabinet is
not directed towards the CMM.
Only for MCC 800: the distance between the back of the controller cabinet and the wall must be
at least 500 mm. When the back is fully open, the distance to the wall must be at least 800 mm.
The air exchange between the controller cabinet and the environment must not be obstructed.
Dimensions
Height [mm]
Width [mm]
Depth [mm]
TCC controller cabinet (standard)
640
600
680
MCC 800 controller cabinet (optional)
800
800
800
MCC 800 controller cabinet (option) with cooling unit (back
door)
800
800
1060