ACCURA Installation site requirements
38
Foundation and floor load
Information about surface pressure
NOTE
The specified surface pressure indicates the maximum load applied to the floor covering. This value
must not be used to calculate the permissible floor or ceiling load. The load conditions at the instal-
lation site vary individually and must be clarified and/or calculated by a structural engineer before in-
stallation of the CMM. See the chapter on installation dimensions for the required information.
ACCURA with active damping
Alignment of air dampers for X1600
The base is comprised of four air dampers and four leveling units which are suitable for screed floors as per DIN
18560. The air dampers and leveling unit are only placed on the foundation (floor) as shown in the illustration.
If the quality specified in DIN 18560 is not guaranteed or the foundation is in poor condition, then the enclosed
spacer plates (1 or 3 mm thick) are required.
700
2440
X1600 / Y2400
520
0
0
1820
Ø 384
X1600 / Y4200
1100
3840
520
0
0
1820
Ø 530
850
2890
X1600 / Y3000
520
0
0
1820
Ø 530
3740
4940
3140
1
Outer edge of granite table
2
Air damper (BiAir3) with leveling unit
3
Air damper (BiAir4) with leveling unit