<6. Troubleshooting>
6-9
IM 12Y18A03-01EN
1st Edition : Oct. 04, 2019-00
Displayed
Event Code
Description
Cause
Troubleshooting
E109
CH1 BAD SLOPE
Last calibration produced a bad slope
on channel 1. Slope is less than -40
mV/decade or more than -75 mV/
decade or the calibration standards
may have been interchanged.
Recalibrate the analyzer using fresh
standards. Recalibrate using the
correct sequence of standards.
E110
CH2 BAD SLOPE
Last calibration produced a bad slope
on channel 2. Slope is less than -40
mV/decade or more than -75 mV/
decade or the calibration standards
may have been interchanged.
Recalibrate the analyzer using fresh
standards. Recalibrate using the
correct sequence of standards.
E113
CH1 CAL OVERDUE Calibration is overdue on channel
1 by more than 50% of the user
specified time interval
Recalibrate the analyzer
E113
CH2 CAL OVERDUE Calibration is overdue on channel
2 by more than 50% of the user
specified time interval
Recalibrate the analyzer
E121
CH1 TEMP BROKEN Temperature sensor on channel 1
broken or out of range
Check ATC cable connections and
probe for damage. Replace if needed.
E122
CH2 TEMP BROKEN Temperature sensor on channel 2
broken or out of range
Check ATC cable connections and
probe for damage. Replace if needed.
E125
CH1 BAD OFFSET
Last calibration produced a bad E
0
offset on channel 1
Recalibrate the analyzer using
fresh standards. Check electrode
connections. Replace electrode, if
needed.
E126
CH2 BAD OFFSET
Last calibration produced a bad E
0
offset on channel 2
Recalibrate the analyzer using
fresh standards. Check electrode
connections. Replace electrode, if
needed.
6.4
Resetting the Analyzer
The WE430 Hydrazine analyzer allows the user to reset the analyzer through the setup mode or
by a hard reset. Resetting the analyzer will restore all settings in the setup mode to factory default
values.
WARNING
Resetting the analyzer will erase all data logs including calibration, and password protection
settings for setup and calibration.
The analyzer will reset setup parameters to factory default settings.
Operator Action
Scrolling Marquee
Main Display
Notes
Press
in the measure modeto enter the
setup mode
Last menu used in setup
mode
SEL
SCrn
SETUP appears as
mode
Press
/
until RSET appears in the
marquee
RSET
(flashing)
SEL
SCrn
Press
to reset the analyzer
PRESS TEST
VIEWENTER TO RESET
rSEt
?