<4. Calibration>
4-5
IM 12Y18A03-01EN
1st Edition : Oct. 04, 2019-00
Step
Operator Action
Scrolling Marquee
Main Display
Notes
20
Press
to accept the new slope
and E
0
values
or
Press
to abort the calibration.
The new slope and E0 values will not
be saved. The last saved calibration
prior to the aborted calibration will be
used in the measure mode.
NEW CALIBRATION
SLOPE AND E
0
PRESS ENTER TO
ACCEPT OR PRESS
EXIT TO ABORT
WITHOUT SAVING
Calculated Slope
and E
0
If
is pressed,
refer to the Calibration
Abort Steps section.
21
Turn off the dynamic calibrator.
Close the standard injection port by
turning the thumbscrew 1/4 of a turn to
the left, so the dot is perpendicular with
the direction of the flow.
Press
.
TURN PUMP OFF
CLOSE INJECTION
VALVE
DOT DOWN
PRESS ENTER
22
System will proceed to measure mode. Actual temperature
reading
Actual
concentration
reading
System enters
measure mode, but
will remain on hold
until the hold timer
expires.
Default hold time is
30 minutes.
4.5
Calibration Error Codes
At any point during a calibration, the appropriate error message will be displayed if there is a
calibration related error.
The following table lists common calibration error codes with their marquee display and required
operator actions.
Step
Operator Action
Scrolling Marquee Main Display
Notes
Stability Time Out Error
Error
41
Press
to continue the calibration
despite the reading instability.
or
Press
to abort the calibration and
refer to Exit 3. Repeat the calibration or
refer to the troubleshooting section.
READINGS WERE
UNSTABLE PRESS
ENTER TO CONTINUE
CALIBRATION OR
PRESS EXIT TO
ABORT
E041
The required system
stability was not
reached and a time out
error occurred.
Perform electrode
cleaning, conditioning
and troubleshooting.
Calibration Errors
Error
42
Press
to abort the calibration and
refer to Exit 3.
Verify that the correct standards were
used during the calibration, the standard
values were entered correctly and in the
right order and the electrodes are working
properly and then repeat the calibration.
INVALID SLOPE
PRESS ENTER
E042
The calculated slope
was invalid. Standard
values were likely
entered in the wrong
sequence or wrong
standard values were
entered.
Error
107
Press
to abort the calibration and
refer to Exit 3.
CALIBRATION
DATA TOO CLOSE
TOGETHER PRESS
ENTER
E107
Use new standards.
Perform electrode
cleaning, conditioning
and troubleshooting.
Error
109
Press
to continue the calibration
despite the bad slope.
or
Press
to abort the calibration and
refer to Exit 3.
CH1 BAD SLOPE
PRESS ENTER
E109
The calculated slope
was outside of the
recommended range.
Perform electrode
cleaning and
conditioning, refer to
the troubleshooting
section and repeat the
calibration.