<3. Analyzer Operation>
3-18
IM 12Y18A03-01EN
1st Edition : Oct. 04, 2019-00
Configuring the analog output mA2
Default
Operator Action
Scrolling Marquee
Main Display
Notes
mA2
mA2 (flashing)
SEL
SCrn
SEL CH1 or SEL CH2
on main display of
two channel analyzer
Press
to set the analog output 2
4-20
Press
/
to scrollbetween
4-20 and 0-20
SELECT 0-20 OR
4-20
4-20 or 0-20
(flashing)
Press
to accept the setting and
advance to the next screen
AOUT 2 SENSOR
OR TEMPERATURE
SEnS
(flashing)
If choosing Sensor:
Press
/
to scroll between
SEnS or tEnP and select SEnS
AOUT 2 SENSOR
OR TEMPERATURE
SEnS
(flashing)
Press
to accept the setting and
advance to the next screen
LOg
Press
/
to scroll between
LOg and LIn
SELECT LOG OR
LINEAR OUTPUT
LOg or LIn (flashing)
Press
to accept the setting and
advance to the next screen
1 ppb
mA SENSOR
OUTPUT LOW
VALUE
LO
###.#
(first digit flashing)
Press
to move the decimal point
Press
/
to set the first digit
Press
to move to the next digit
Press
/
to set the values of
the remaining digits and press
to move through the remaining digits
mA SENSOR
OUTPUT LOW
VALUE
LO
###.#
(change flashing
digit)
Press
to accept the setting and
advance to the next screen
100 ppb
mA SENSOR
OUTPUT HIGH
VALUE
HI
###.#
(first digit flashing)
Press
to move the decimal point
Press
/
to set the first digit
Press
to move to the next digit
Press
/
to set the values of
the remaining digits and press
to move through the remaining digits
mA SENSOR
OUTPUT HIGH
VALUE
HI
###.#
(change flashing
digit)
Press
to accept the setting and
advance to the next screen
CH1 or CH2
(flashing)
SEL
SCrn
Displayed for two
channel analyzer
Press
to return to the channel
specific menu options in the setup
mode
Action required for
two channel analyzer
mA2 (flashing)
SEL
SCrn
SEL CH1 or SEL CH2
on main display of
two channel analyzer
Press
to scroll to the next menu
If choosing temperature:
Press
/
to scroll between
SEnS or tEnP and select tEnP
AOUT 2 SENSOR
OR TEMPERATURE
tEnP (flashing)
Press
to accept the setting and
advance to the next screen
05.0 C
mA TEMP
OUTPUT LOW
VALUE
LO
##.# c
(first digit flashing)