<3. Analyzer Operation>
3-10
IM 12Y18A03-01EN
1st Edition : Oct. 04, 2019-00
Default
Operator Action
Scrolling Marquee
Main Display
Notes
TIME
TIME
(flashing)
SEL
SCrn
Press
to set the time
00:01
ENTER 24HR TIME
HOUR/MINUTE
##:## (hour : minute)
(first digit flashing)
Press
/
to set the first digit
Press
to move to the next digit
Press
/
to set the valuesof
the remaining digits and press
to move through the remaining digits
ENTER 24HR TIME
HOUR/ MINUTE
##:## (hour : minute)
(change flashing digit)
Set in 24hour time
format
Press
to accept the time setting
and return to the main setup mode
TIME
(flashing)
SEL
SCrn
Press
to scroll to the next menu
Default
Operator Action
Scrolling Marquee
Main Display
Notes
LOG
LOG
(flashing)
SEL
SCrn
Press
to set the log interval
00:00
SET LOG TIME IN
HOUR/ MINUTE
##:## (hour : minute)
LOg
(first digit flashing)
Press
/
to set the first digit
Press
to move to the next digit
Press
/
to set the values of
the remaining digits and press
to move through the remaining digits
SET LOG TIME IN
HOUR/ MINUTE
##:## (hour : minute)
LOg
(change flashing
digit)
To disable the log
enter 0000 for the log
interval
The minimum log
interval is 1 minute
and the maximum log
interval is 99 hours
and 59 minutes
Press
to accept the log setting
and return to the main setup mode
LOG
(flashing)
SEL
SCrn
Press
to scroll to the next menu
WARNING
The reset command will erase all operator settings, logs and calibration data. The analyzer will
need to be set up and calibrated again before it can resume operation.