<4. Calibration>
4-8
IM 12Y18A03-01EN
1st Edition : Oct. 04, 2019-00
Step
Operator Action
Scrolling Marquee
Main Display
Notes
5
Press
/
to select the
channel that will be calibrated and
press
to begin the calibration.
CH1 or CH2
(flashing)
CAL CHnL
Action required for
two channel analyzer
6
DKA or OFFL
(flashing)
CAL tYPE
7
Press
/
to select OFFL and
press
to accept the setting
OFFL
(flashing)
CAL tYPE
8
Wait 1 to 10 minutes.
MEASURING
SAMPLE WAIT 1
MINUTE AWAITING
DATA STABILITY
Actual readings
flashing
System continuously
evaluates the
data until a stable
condition is met.
9
Take a QC sample from the 2118XP
analyzer bypass for laboratory
analysis.
Press
.
TAKE SAMPLE
FROM BYPASS
DRAIN PRESS
ENTER TO USE
METER THEN
PRESS CAL
TO ENTER LAB
RESULTS
LAb
10 Perform laboratory analysis on QC
sample and return with results.
Actual temperature
reading
Actual concentration
reading
System will return to
measure mode and
operate as normal.
CAL and MEASURE
appear as mode.
11 Press
to resume the calibration.
HOLD and CAL
appear as mode.
12 If the calibration password is active,
enter the calibration password as
described in step 3.
Press
to accept the password.
ENTER PASSWORD PASS
# # # #
13
CH1 or CH2
(flashing)
CAL CHnL
Displayed for two
channel analyzer.
14 Press
/
to select the
channel that will be calibrated and
press
to begin the calibration.
CH1 or CH2
(flashing)
CAL CHnL
Action required for
two channel analyzer.
15 Press
/
to select OFFL and
press
to accept the setting.
OFFL
(flashing)
CAL tYPE
16 Enter the concentration value of the
QC sample obtained from laboratory
analysis:
Press
to move the decimal point
and set the value as ppm or ppb.
Press
/
to set the first digit.
Press
to move to the next digit.
Press
/
to set the values of
the remaining digits and press
to move through the remaining digits.
Press
to save the value.
LAB ANALYSIS
READY INPUT
CONC VALUE FROM
LAB PRESS ENTER
TO CONTINUE OR
PRESS EXIT TO
ABORT
COnC
# # # . #
(change flashing
digit)
17
NEW CALIBRATION
E
0
PRESS ENTER
TO ACCEPT OR
PRESS EXIT TO
ABORT
Actual E
0
value
The system
will display the
recalculated the E
0
.
18 Press
.
The system willreturn
to the measure mode.
Only MEASURE
should appear as
mode.