<2. Analyzer Preparation>
2-6
IM 12Y18A03-01EN
1st Edition : Oct. 04, 2019-00
2.3.1
Terminal Assignments
Terminal Layout
Terminal Layout
Terminal Layout
Terminal Layout
1 mA1 output
9 Relay 1
26 Sensing electrode 28 Do not connect
2 GND common ground
10 Relay 1
27 Do not connect
29 Do not connect
3 mA2 output
11 Relay 2
30 Preamp power
4 Air pump (ISE only)
12 Relay 2
31 Preamp ground
5 Air pump (ISE only)
13 Relay 3
32 Shield
6 Shield ground for
conductivity
14 Relay 3
33 Shield
7 Do not connect
15 Do not connect
34 Jumper to pin 26
when using preamp
8 Do not connect
16 Temperature ground
17 Temperature drive
18 Temperature sense
19 Solution ground
20 Conductivity drive +
21 Conductivity sense +
22 Conductivity sense -
23 Conductivity drive -
24 Reference electrode
25 Jumper to pin 24
when using preamp
9
10 11 12 13 14 15 16 17
1
2
3
4
5
6
7
8
28 29 30 31 32 33 34
26 27
18 19 20 21 22 23 24 25
Figure 2.6 Terminal Assignments
2.3.2
Electrode Wiring Assignments
Iodide Electrode
26
Sensing electrode
Connect clear wire
33
Shield
Connect black wire
Reference Electrode
24
Reference electrode
Connect clear wire
32
Shield
Connect black wire
2.4
Installation of Reagent and Diffusion Tubing
WARNING
The reagent is hazardous. Use protective glasses and gloves. Refer to the bottle label for
precautions and work in a fume hood or well-ventilated area. Avoid contact with skin or clothes. In
case of skin contact, flush skin immediately with water to prevent burns.