<5. Analyzer Maintenance>
5-2
IM 12Y18A03-01EN
1st Edition : Oct. 04, 2019-00
5.4
Monthly Maintenance
Calibration frequency is operator dependent for the most accurate and precise measurements.
5.4.1 Calibration
• Replacement of the reagent requires calibration. Refer to Section 4.4, Performing a DYN
Calibration.
• Calibration intervals can be programmed using the setup mode. This will prompt the
operator to recalibrate after a specific amount of time has passed. Refer to Section 3.1, Use
of the Setup Mode.
5.4.2
Replacement of Sample Inlet Filter
1. If the sample flow can be adjusted with the pressure regulator to 50 mL/minute or more, skip
this procedure.
2. Turn off sample flow by turning the inlet valve to the right.
3. Remove the filter hex cap using a 1 inch open-ended wrench.
4. Separate the filter cap with bypass/needle valve assembly from the filter body, retaining the
ring.
5. Remove the filter element and soft gasket (if necessary).
6. Replace the gasket and press a new sintered element into the filter body.
7. Replace the hex cap and bypass/needle valve assembly and reposition the bypass drain
line into the drain assembly.
8. Retighten the hex cap to approximately 75 pound-inch in torque.
9. Turn on the sample flow by turning the inlet valve left.
10. Check for leaks.
5.4.3
Polish Iodide Electrode
Refer to Section 2.6, Installation of a New Iodide Electrode.
Polish the iodide electrode according to the instructions that are included with the polishing strips.
Rinse the electrode thoroughly with distilled water, insert the iodide electrode into its holder in the
flow cell and then plug the electrode cable marked “Sensing” into the top of the electrode.
5.4.4
Air Pump Check
The reference electrode fill solution level should drop at least 1/8’’ per month, which indicates
good flow through the electrode frit. If the flow rate is low, the air pump should be tested for
proper operation.
To test the air pump, remove the pushpin and tubing from the reference reservoir. Place the tip of
the pushpin in a beaker of water. Vigorous bubbling should be observed. Normally, the air pump
delivers at least 500 cc/minute of air. Replace the air pump if weak bubbling or no bubbling is
observed. To ensure a tight seal on the reservoir, replace it with a fresh bottle of fill solution (Cat.
No. 181073).
5.4.5
Replacement of Reagent and Diffusion Tubing
Assembly
WARNING
The reagent is hazardous. Use protective glasses and gloves and work in a well-ventilated area
or a fume hood. Avoid contact with skin or clothes. In case of skin contact, flush skin immediately
with water to prevent burns. Refer to the bottle label for precautions.