<3. Analyzer Operation>
3-13
IM 12Y18A03-01EN
1st Edition : Oct. 04, 2019-00
NOTE
Changing the application mode will reset the analyzer, restore the default parameters and clear
any saved calibration information.
Default
Operator Action
Scrolling Marquee
Main Display
Notes
MEAS
MEAS
(flashing)
SEL
SCrn
SEL CH1 or SEL CH2
on main display of
two channel analyzer
Press
to set measure
parameters
3
SET NUMBER
OF SIGNIFICANT
DIGITS
SIg 3
(flashing)
Press
/
to scroll through
2, 3 and 4
SET NUMBER
OF SIGNIFICANT
DIGITS
SIg 2, 3 or 4
(flashing)
Press
to accept the setting and
advance to the next screen
OFF
SHOW MV ON
SINGLE CHANNEL
DISPLAY
OFF
(flashing)
Press
/
to scroll between
OFF and On
SHOW MV ON
SINGLE CHANNEL
DISPLAY
OFF or On (flashing)
Press
to accept the setting and
advance to the next screen
AUtO
SELECT ISE UNIT UnIt AUtO
(flashing)
Press
/
to scroll through
AUtO, PPb and PP
SELECT ISE UNIT UnIt
AUtO, PPb or PP
(flashing)
PP is used as an
abbreviation for ppm
Press
to accept the setting and
return to the main setup mode
CH1 or CH2
(flashing)
SEL
SCrn
Displayed for two
channel analyzer
Press
to return to the channel
specific menu options in the setup
mode
Action required for
two channel analyzer
MEAS
(flashing)
SEL
SCrn
SEL CH1 or SEL CH2
on main display of
two channel analyzer
Press
to scroll to the next menu