<3. Analyzer Operation>
3-17
IM 12Y18A03-01EN
1st Edition : Oct. 04, 2019-00
Configuring the analog output mA1
Default
Operator Action
Scrolling Marquee
Main Display
Notes
mA1
mA1
(flashing)
SEL
SCrn
SEL CH1 or SEL
CH2 on main display
of two channel
analyzer
Press
to set the analog output
1
4-20
Press
/
to scroll between
4-20 and 0-20
SELECT 0-20 OR
4-20
4-20 or 0-20
(flashing)
Press
to accept the setting and
advance to the next screen
Press
/
to scroll between
LOg and LIn
SELECT LOG OR
LINEAR OUTPUT
SOUt
LOg or LIn
(flashing)
Press
to accept the setting and
advance to the next screen
1 ppb
mA SENSOR
OUTPUT
LOW VALUE
LO
###.#
(first digit flashing)
Press
to move the decimal
point
Press
/
to set the first digit
Press
to move to the next digit
Press
/
to set the values
of the remaining digits and press
to move through the remaining
digits
mA SENSOR
OUTPUT LOW
VALUE
LO
###.#
(change flashing
digit)
Press
to accept the setting and
advance to the next screen
100 ppb
mA SENSOR
OUTPUT HIGH
VALUE
HI
###.#
(first digit flashing)
Press
to move the decimal
point
Press
/
to set the first digit
Press
to move to the next digit
Press
/
to set the values
of the remaining digits and press
to move through the remaining
digits
mA SENSOR
OUTPUT HIGH
VALUE
HI
###.#
(change flashing
digit)
Press
to accept the setting and
return to the main setup mode
CH1 or CH2
(flashing)
SEL
SCrn
Displayed for two
channel analyzer
Press
to return to the channel
specific menu options in the setup
mode
Action required for
two channel analyzer
mA1 (flashing)
SEL
SCrn
SEL CH1 or SEL
CH2 on main display
of two channel
analyzer
Press
to scroll to the next menu