<3. Analyzer Operation>
3-5
IM 12Y18A03-01EN
1st Edition : Oct. 04, 2019-00
l
DISP
Set the automatic lighting options for the backlit display:
• AUtO – Brightness will change in response to ambient light source
• On – Backlit display is always on
• OFF – Backlit display is always off
• Default display – AUtO
l
MDL
Sets the analyzer to the hydrazine application mode.
NOTE
Changing the application mode will reset the analyzer, restore the default parameters and clear
any saved calibration information.
l
MEAS
Set the number of significant digits, mV display option and concentration units displayed when in
the measure mode:
• Set the number of significant digits displayed measure mode
• Scroll through 2, 3 or 4 significant digits
• Default significant digits – 3
• Enable mV values to be displayed on the second line
• Scroll between On or OFF
• Default mV setting – OFF
• Set the displayed concentration units
• Scroll through AUtO (automatically ranges from ppb to ppm), PPb (parts per billion,
ppb), or PP (parts per million, ppm)
• Default displayed concentration units – AUtO
l
HOLD
Set the time that the system will remain on hold before the actual measurements are displayed
after a calibration:
• Once the hold time expires, the system implements any programmed changes to settings in
the setup mode
• After a calibration, the hold function allows the operator to rinse the electrodes prior to
recording actual measurement values
• Default hold time – 30 minutes
l
TADJ
Adjust the temperature reading from the ATC probe by ± 5.0 °C:
• Default adjustment – 0.0 °C