<4. Calibration>
4-1
IM 12Y18A03-01EN
1st Edition : Oct. 04, 2019-00
4. Calibration
4.1
Calibration Setup
The WE430 Hydrazine analyzer provides simple and fast calibration cycles using a dynamic two
point calibration procedure, which requires the use of the dynamic calibrator.
To perform a dynamic two point calibration, the WE430 analyzer is first connected to a reductant-
free sample stream. An appropriate diluted standard is prepared and then a syringe is filled
with the diluted standard and mounted on the calibrator. The calibrator is then mounted near
the analyzer and the syringe tubing is connected to the standard injection port. By adjusting the
calibrator pump settings, two different flow rates produce two known standards that are diluted
into the sample background.
Prior to performing a calibration, refer to Section 3.1, Use of the Setup Mode for instructions on
how to properly configure the WE430 analyzer and set the analyzer to the hydrazine application
mode.
n
DYN Calibration Recommendations
When the WE430 Hydrazine analyzer is first commissioned, it is advisable to flush out fluidics
system overnight prior to the initial calibration and use. The electronics need not be turned on at
this time. Refer to Chapter 4, Rinsing the Flow Cell.
• Install the sensing electrode, reference electrode and ATC probe in their respective flow cell
holders. Refer to Section 2.6, Installation of a New Iodide Electrode, Section 2.7, Installation
of ATC Probe and Section 2.8, Installation of a New Reference Electrode.
• To ensure the analyzer achieves stability and to minimize interference during calibration, the
operator should minimize activity on the system during calibration.
• Maximum system accuracy is ensured through a fast, easy and accurate calibration
performed in the expected sample range. This calibration procedure uses equipment
supplied in the calibration kit including the dynamic calibrator.
• Wait at least one hour to calibrate the analyzer after changing the reagent.
• Use an oxygen scavenger-free sample during calibration.
• Fill and empty the syringe with standard at least three times before connecting the syringe to
the dynamic calibrator.
• Always use a new syringe for each calibration.