12-17
IM 04P01B01-01E
Specifications
1
2
3
4
5
6
7
8
9
10
11
12
App
Index
24 VDC/AC Power Supply Operation (/P1)
Item
Specifications
Rated supply voltage
24 VDC/AC
Allowable power supply voltage range
21.6 to 26.4 VDC/AC
Withstand voltage
1000 VAC at 50/60 Hz for one minute (between the power terminal and the ground terminal)
Rated power supply frequency
50/60 Hz (for AC operation)
Allowable power supply frequency range
50 Hz ± 2% or 60 Hz ± 2% (for AC operation)
Effects of power supply voltage fluctuation
The fluctuation in the measured values is within ±1 digit and that in the recording is within ±0.1% in
the recording span for the 21.6 to 26.4 VAC/DC range.
Effects of power supply frequency fluctuation
The fluctuation in the measured values and recording is ±(0.1% of rdg + 1 digit) at the rated
frequency ±2 Hz.
Rated power
25 VA (for DC) or 35 VA (for AC)
Power consumption
Model
24 VDC
24 VAC
Maximum
1 to 4 pen models
Approx. 7 VA*
Approx. 13 VA*
Approx. 35 VA
6 dot model
Approx. 8 VA*
Approx. 13 VA*
Approx. 35 VA
* When balanced
Calibration Correction (/CC1)
Item
Specifications
Number of calibration points 2 to 16 (can be specified for each channel)
Correction mode
Revise or absolute value (can be specified for each channel)
Correctable range
DC input (DC voltage, TC, and RTD) and linear scaling (DC voltage, TC, RTD, and 1-5V)
(Not allowed on ON/OFF input (including linear scaling), delta computation, square root
computation, and computation channels)
Correction range
• (1
st
correction point + 1
st
correction value) < (2
nd
correction point + 2
nd
correction value) < ...
• Corrected value (revise value: correction point + correction value, absolute value: correction
value) within the measurable range and scaling range (–5% to 105% or –20000 to 30000) of each
input range
• Setting limits
Left span (0%) < Right span (100%)
Left scale (0%) < Right scale (100%)
Others
Calibration correction and bias functions cannot be used simultaneously (common to all channels)
Header Printout (/BT1)
Item
Specifications
Printout function
When recording starts or stops, prints out the batch name, comment, date/time, and chart speed.
Messages can also be printed out including measured values*.
Printed items
Recording start/recording end printouts
Batch name: Batch number (up to 26 characters) and lot number (4 or 6 digit number, or Off)
Comment (up to 32 characters x 5 lines)
Chart speed
Date/time: Year, month, date, time, minute, second (date format depends on the date printout/
display format)
Printout On/Off selection: Batch name, chart speed, and date/time
Message printouts
In the 5 message formats, measured values*, strings (up to 16 characters), and date/time can be
arbitrarily combined and printed out (up to 35 characters total).
* Computed values can also be included on models with the /M1 option.
SD Memory Card (/EM1)
Item
Specifications
Measured data save
Saves measured data to the SD memory card. File is divided every about 30 MB. .RXE extension.
Measured data replay
Universal Viewer (software) for viewing and printing.
Data saved with a dot model to an SD memory card can be replayed on chart paper using a
recorder of the same model with equivalent or more functions.
Setting parameters save
and load
Saves recorder setup data to an SD memory card. .PUL extension.
Loads setup data from an SD memory card into the recorder.
12.5 Specifications of Optional Functions