IM 04P01B01-01E
5-2
5.1 Setting the Input Range
Description
• Selectable Range of Input Range, Span Left, and Span Right
The input range, span left, and span right can be set in the range shown below. Span
left and span right cannot be set to the same value.
DC voltage (Volt)
Range Type
Selectable Span Range
20 mV
–20.00 to 20.00 mV
60 mV
–60.00 to 60.00 mV
200 mV
–200.0 to 200.0 mV
2 V
–2.000 to 2.000 V
6 V
–6.000 to 6.000 V
20 V
–20.00 to 20.00 V
50 V
–50.00 to 50.00 V
Thermocouple (TC)
Range Type
Selectable Span Range (°C) Selectable Span Range (°F)
R
0.0 to 1760.0°C
32 to 3200°F
S
0.0 to 1760.0°C
32 to 3200°F
B
0.0 to 1820.0°C
32 to 3308°F
K
–200.0 to 1370.0°C
–328 to 2498°F
E
–200.0 to 800.0°C
–328.0 to 1472.0°F
J
–200.0 to 1100.0°C
–328.0 to 2012.0°F
T
–200.0 to 400.0°C
–328.0 to 752.0°F
N
0.0 to 1300.0°C
32 to 2372°F
W
0.0 to 2315.0°C
32 to 4199°F
L
–200.0 to 900.0°C
–328.0 to 1652.0°F
U
–200.0 to 400.0°C
–328.0 to 752.0°F
WRe
0.0 to 2400.0°C
32 to 4352°F
RTD
Range Type
Selectable Span Range (°C) Selectable Span Range (°F)
PT (Pt100)
–200.0 to 600.0°C
–328.0 to 1112.0°F
JPT (JPt100)
–200.0 to 550.0°C
–328.0 to 1022.0°F
RTD
Range Type
Selectable Span Range (°C) Selectable Span Range (°F)
PT (Pt100)
–200.0 to 600.0°C
–328.0 to 1112.0°F
JPT (JPt100)
–200.0 to 550.0°C
–328.0 to 1022.0°F
/N1 Option
/N3 Option
Selections
Input Type
Selections
Input Type
Note
Cu1
Cu10 (GE)
PR
PR40-20
Thermo
couple
Cu2
Cu10 (L&N)
PLATI
PLATINEL
Cu3
Cu10 (WEED)
NiMo
NiNiMo
Cu4
Cu10 (BAILEY)
W/WRe
W/WRe26
Cu5
Cu10: a = 0.00392 at 20°C
N2
Type N (AWG14)
Cu6
Cu10: a = 0.00393 at 20°C
Kp
Kp vs Au7Fe
Cu25
Cu25*: a = 0.00425 at 0°C
Pt4
Pt25
RTD
Pt3
Pt50
Ni1
Ni100 (SAMA)
Ni2
Ni100 (DIN)
Ni3
Ni120
J263
J263*B
Cu8
Cu53
Cu9
Cu100: a = 0.00425 at 0°C
Summary of Contents for MuR10000
Page 2: ......
Page 14: ...Blank...
Page 18: ...Blank...
Page 58: ...Blank...
Page 72: ...Blank...
Page 92: ...Blank...
Page 220: ...Blank...
Page 268: ...Blank...