• Computation Start/Stop (/M1 Option)
• Remote input signal: Rising edge signifies start; falling edge signifies stop
• Starts/stops the computation.
• If the computation is started, applying a rising edge signal produces no effect. If the
computation is stopped, applying a falling edge signal produces no effect.
• Computation Reset (/M1 Option)
• Remote input signal: Trigger
• The computed results of computation channels are reset. This is valid only while
the computation is stopped. For all other cases, applying the remote signal
produces no effect.
• Priority to Remote Recording (/BT1 Option)
• Remote input signal: Edge (rising/start or falling/stop)
• Starts/stops recording.
• When started with a remote signal (on a remote signal rise), stop per key operation
or communication is disabled.
• Switching Batch Comment (/BT1 Option)
• Remote input signal: Level
• Switches between Start printout and Start printout 2, and between End printout
and End printout 2 depending on the status of the batch comment switching signal
when recording is started/stopped remotely.
• When starting/stopping by key operation, performs Start printout and End printout.
• Starting and Stopping Data Saving (/EM1 Option)
• Remote input signal: Edge (rising/start or falling/stop)
• Starts or stops data saving.
• If data saving is in progress, applying a rising edge signal produces no effect. If
data saving is not in progress, applying a falling edge signal produces no effect.
Remote Signal (Edge, Trigger, and Level)
The above actions are carried out on the rising or falling edge of the remote signal (edge),
the ON signal lasting at least 250 ms (trigger), or the ON/OFF signal (level).
Rising/Falling edge
Trigger
Level
Rising
Falling
250 ms or more
Operates at the
secondary chart speed
250 ms or more
For contact inputs, the remote signal rises when the contact switches from open to closed
and falls when the contact switches from closed to open. For open collector signals, the
remote signal rises when the collector signal (voltage level of the remote control terminal)
goes from high to low and falls when the collector signal goes low to high.
<Related Topics>
Assigning functions to the remote control input terminals: Section 7.18
Setting the secondary chart speed: Section 6.9
1.8 Remote Control Function (/R1 Option)