<2.
Installation, Wiring and Piping>
2-3
IM 11T03E01-51E
2.1.3
GD40V (CSA Explosion-proof and Intrinsically Safe
Approval)
Sealing Fitting
Conduit
500 mm (20") Max.
NON-HAZARDOUS
LOCATIONS
HAZARDOUS
LOCATIONS
GD40V
F0203.ai
Figure 2.2
1. GD40V Gas density detector for use in hazardous (classified) locations:
• Explosion-proof for Class I, Division 1, Groups B, C and D;
Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G with Intrinsically
Safe sensor for Class I, II, III, Division 1, Groups B, C, D, E, F and G.
• Enclosure Rating:
Type 4X
• Temperature Code:
T5
• Ambient Temperature: -10 to 60°C
2. Wiring
• All wiring shall comply with Canadian Electrical and Local Electrical codes.
• Note a warning label worded as follows.
WARNING
A seal shall be installed within 500 mm (20 inches) of the enclosure. Refer to Figure 2.2.
3. Operation
• Note a warning label worded as follows.
WARNING
• Open circuit before removing cover.
• Substitution of components may impair intrinsic safety.
• Take care not to generate mechanical spark when access to the instrument and
peripheral devices in hazardous locations.
4. Maintenance and Repair
•
The instrument modification or parts replacement by other than authorized representative
of Yokogawa Electric Corporation is prohibited and will void the certification of CSA
International.