<1. Specifications>
1-3
IM 11T03E01-51E
CAUTION
This instrument is a Class A product, and it is designed for use in the industrial environment.
Please use this instrument in the industrial environment only.
l
Characteristics
GD402 specification list
Item
Density kg/m
3
Density lb/ft
3
Specific Gravity
Molecular Weight Concentration vol%
Range
0 - 6 (compensated)
0 - 60 (physical)
0 - 0.4 (compensated)
0 - 4 (physical)
0 - 5
0 - 140
0 - 100
Minimum Range 0.1
0.01
0.1
4
Concentration equivalent
to 100 kg/m
3
Response Time 90% approx. 5 sec
approx. 5 sec
approx. 5 sec
approx. 5 sec
approx. 5 sec
Linearity
+/-1% FS
+/-1% FS
+/-1% FS
+/-1% FS
+/- 1
Repeatability
+/- 0.001
or +/-0.5%FS *
+/- 0.0001
or +/-0.5%FS*
+/- 0.001
or +/-0.5%FS*
+/- 0.02
or +/-0.5%FS*
+/- 0.5% or Concentration
equivalent to +/-0.001 kg/m
3
*
Long term
stability
+/- 0.003/month +/- 0.002/month
+/- 0.003/month +/- 0.07/month
Concentration equivalent
to +/- 0.003 kg/m
3
/month
*:
Whichever is greater
Density is the basic measurement, the other representations are derived from the Density data.
Item
H
2
in Air vol% H
2
in CO
2
vol% Air in CO
2
vol%
Caloric value MJ/m
3
British Thermal Unit KBTU/ft
3
Range
85 - 100
0 - 100
0 - 100
0 - 130
0 - 3.5
Minimum
Range
–
–
–
Caloric value equivalent
to 0.100 kg/m
3
Caloric value equivalent
to 0.100 kg/m
3
Response
Time 90%
approx. 5 sec approx. 5 sec approx. 5 sec
approx. 5 sec
approx. 5 sec
Linearity
+/- 1
+/- 1
+/- 1
+/- 1%FS
+/- 1%FS
Repeatability
+/- 0.5
+/- 0.5
+/- 0.5
+/- 0.5%FS or Caloric value
equivalent to 0.001kg/m
3
*
+/- 0.5%FS or Caloric value
equivalent to 0.001 kg/m
3
*
Drift
+/- 0.5/month +/- 0.5/month +/- 0.5/month
Caloric value equivalent
to +/- 0.003 kg/m
3
/month
Caloric value equivalent
to +/- 0.0025/month
Caloric Value and BTU are possible representations of the Density.
GD402 does not contain table information, only a single mathematical equation.
*: Whichever is greater
1.1.2
GD40G, T, V, R Detector
l
Material exposed to gas
SUS 316 stainless steel, Acrylonitrile Butadiene Rubber and Fluorine-contained Rubber
(O-ring)
l
Ambient conditions
Temperature:
-10 to 60°C (14 to 140 °F)
Humidity:
5 to 95%RH
Installation: Pipe-mounted or on panel
Construction:
Explosion-proof and Intrinsically safe. (See 1.1.1 “System Components”)
Though the detector construction makes it relatively insensitive to sudden changes in
the gas temperature, extra precision can be achieved by keeping ambient temperature
conditions as constant as possible. In measurements where optimum precision
is required it is therefore not recommended to install the detector in an outdoors
environment, especially not if such installation is prone to direct sunlight.
l
Finish
Cover:
Equivalent to Munsell 0.6GY3.1/2.0
Case:
Equivalent to Munsell 2.5Y8.4/1.2
Summary of Contents for GD402G /M1
Page 21: ...Blank Page ...
Page 37: ...Blank Page ...
Page 81: ...Blank Page ...
Page 115: ...Blank Page ...
Page 159: ...Blank Page ...
Page 167: ......
Page 169: ......
Page 171: ...Blank Page ...