<1. Specifications>
1-1
IM 11T03E01-51E
1. SPECIFICATIONS
1.1
Specifications
The specifications of the GD402 gas density meter as described below.
Certificate is approved for model GD402G, GD40G.
1.1.1
General Specifications
l
System Components
(1) GD40G, T, V, R Detector: Rainproof for outdoor use (equivalent to IP65/NEMA4X)
(see note under “1.1.2 Ambient conditions” on page 1-5.)
Ambient Temperature :
-10 to 60°C
Ambient Humidity :
5 to 95%RH
GD40G :
General purpose detector. (Non-Explosion-proof)
Electrical connection : 1/2 NPT female
Process connection : 1/4 NPT female
GD40T :
FM Explosion-proof and Intrinsically safe Approval.
Explosion-proof for Class I, Division 1, Groups B, C and D;
Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G with
Intrinsically Safe sensor for Class I, II, III, Division 1, Groups B, C, D, E, F
and G.
Applicable standards:
FM Class 3600:1998, FM Class 3610:2010, FM Class 3615:1989, FM Class
3810:2005, ANSI/ISA 60079-0:2009, ANSI/UL 60079-11:2009, ANSI/NEMA 250:1991
Enclosure :
NEMA Type 4X
Temperature Code :
T5
Electrical connection : 1/2 NPT female
Process connection : 1/4 NPT female
GD40V :
CSA Explosion-proof and Intrinsically safe Approval.
Explosion-proof for Class I, Division 1, Groups B, C and D;
Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G with Intrinsically
Safe sensor for Class I, II, III, Division 1, Groups B, C, D, E, F and G.
Applicable standards:
CAN/CSA-C22.2 No. 25-1966, CAN/CSA-C22.2 No. 30-M1986, CAN/CSA-C22.2
No. 94.2-15, CAN/CSA-C22.2 No. 157-92, CAN/CSA-C22.2 No. 1010.1-92
Enclosure :
Type 4X
Temperature Code :
T5
Electrical connection : 1/2 NPT female
Process connection : 1/4 NPT female
GD40R :
TIIS Explosion-proof and Intrinsically safe Approval.
Explosion-proof code : Exd [ia] IIB+H
2
T5
Temperature Code :
T5
Electrical connection : G1/2 female
Process connection : Rc1/4
(2) GD402G Converter : Rainproof for outdoor use (equivalent to IP65 / NEMA 4X)
Summary of Contents for GD402G /M1
Page 21: ...Blank Page ...
Page 37: ...Blank Page ...
Page 81: ...Blank Page ...
Page 115: ...Blank Page ...
Page 159: ...Blank Page ...
Page 167: ......