<7. HYDROGEN PURITY / PARAMETER SETTING>
7-21
IM 11T03E01-51E
CODE 11: Setting for Compensated Density Measurement
This setting defines the reference temperature and pressure required to obtain the compensated
density.
(a) Setting of reference temperature
[
*C.D.TMP
]
Configurable range:
-20.0 to 80.0 (for
ºC)
-4.0 to 176.0 (for °F )
(b) Setting of reference pressure
[
*C.D.PRS
]
Configurable range:
0.10 to 999.99 (for kPa),
0.0001 to 9.9999 (for MPa),
0.015 to 99.999 (for psi )
[ENT]
11
*CODE
*SERVC
XX.X
*C.D.TMP
XXX.XX
*C.D.PRS
[ENT]
[ENT]
[YES]
F0714.ai
*
1
*
2
*
5
*1:
See "Figure 7.4 Process at the Setting Level" on page 7-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER.]
Press the [YES] or [NO] key to re-set the value.
*4: The position of the decimal point changes depending on the setting of CODE 23.
If the code is set to "0" (°C), the reading is XX.X;
if the code is set to "1" (°F), the reading is XXX.X.
*5: The position of the decimal point changes depending on the setting of CODE 20.
If the code is set to "0" (kPa), the reading is XXX.XX;
if the code is set to "1" (MPa), the reading is X.XXXX;
if the code is set to "2" (psi), the reading is XX.XXX.
*
3
*
4
*
3
Figure 7.14
Setting for Compensated Density Measurement
Summary of Contents for GD402G /M1
Page 21: ...Blank Page ...
Page 37: ...Blank Page ...
Page 81: ...Blank Page ...
Page 115: ...Blank Page ...
Page 159: ...Blank Page ...
Page 167: ......
Page 169: ......
Page 171: ...Blank Page ...