<5. DENSITY / PARAMETER SETTING>
5-19
IM 11T03E01-51E
CODE 05: Setting of Contact Outputs Status
[
*CNTCT
]
This setting defines the status of the contact outputs.
Configurable range: 00 to 15
05
*CODE
*SERVC
XX
*CNTCT
[YES]
F0511.ai
*1:
See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
[ENT]
[ENT]
*
3
*
2
*
1
Figure 5.10
Setting of Contact Outputs Status
Each contact output takes either of the following two status (in table), depending on the value you
set.
Value
ZERO contact
SPAN contact MAINT contact
ALM contact
0
NO
NO
NO
NO
1
NO
NO
NO
NC
2
NO
NO
NC
NO
3
NO
NO
NC
NC
4
NO
NC
NO
NO
5
NO
NC
NO
NC
6
NO
NC
NC
NO
7
NO
NC
NC
NC
8
NC
NO
NO
NO
9
NC
NO
NO
NC
10
NC
NO
NC
NO
11
NC
NO
NC
NC
12
NC
NC
NO
NO
13
NC
NC
NO
NC
14
NC
NC
NC
NO
15
NC
NC
NC
NC
NO: means “normally open”
NC: means “normally closed”