<1. Specifications>
1-2
IM 11T03E01-51E
Ambient Temperature :
-10 to 55°C
Ambient Humidity :
5 to 95%RH
General purpose converter. (Non-Explosion-proof).
Electrical connection:
21mm (0.9inch) in diameter. Pg 13.5 cable glands included
(3) EJX310A Absolute pressure transmitter (optional)
See GS 01C25D01-01EN for EJX310A.
l
Sample Gas conditions
Sample gas: All gases except for corrosive gas and acetylene gas.
Temperature: -10 to 60°C (non-condensing)
Pressure:
Max. 588.4 kPa (abs)
Gas flow:
0.1 to 1 L/min
l
Output Signals
Output 1:
4-20 mA DC
Isolated from inputs; load resistance: 600 Ω maximum (Load resistance of 250-
550 Ω required when in the BRAIN communication mode.)
Output 2:
4-20 mA DC
Isolated from inputs; load resistance: 600 Ω maximum
l
Power Supply
Rated voltage range:
100 to 240 V AC, 24 V DC
Allowable voltage range:
85 to 264 V AC, 21.6 to 26.4 V DC
Rated frequency:
50 or 60 Hz
Allowable frequency range: 47 to 63 Hz
l
Power Consumption
Approximately 12 W.
l
Fuse
250 V 1 A time lag type authorized VDE/SEMKO 100 to 240 V AC Model
250 V 2 A time lag type authorized VDE/SEMKO 24V DC Model
l
Safety, EMC, and RoHS conformity standards
Installation Altitude: 2000 m or less
Category based on IEC 61010-1: II (Note)
Pollution degree based on IEC 61010-1: 2 (Note)
Note: Installation category, called over-voltage category, specifies impulse withstand voltage.
Category II is for electrical equipment.
Pollution degree indicates the degree of existence of solid, liquid, gas or other inclusions which may reduce dielectric
strength. Degree 2 is the normal indoor environment.
Safety Standards:
EN 61010-1, EN 61010-2-030
*: Applied only when GD402G converter is used with GD40G detector.
EMC Standards:
EN 61326-1* Class A Table 2,
EN 61326-2-3, EN 61000-3-2, EN 61000-3-3
EMC Regulatory Arrangement in Australia and New Zealand (RCM) EN61326-1 Class A
Korea Electromagnetic Conformity Standard**
* : Influence of immunity environment (Criteria A): ±50% of F.S.
**: Applied only when GD402G converter is used with GD40G detector.
RoHS:
EN IEC 63000
Summary of Contents for GD402G /M1
Page 21: ...Blank Page ...
Page 37: ...Blank Page ...
Page 81: ...Blank Page ...
Page 115: ...Blank Page ...
Page 159: ...Blank Page ...
Page 167: ......
Page 169: ......
Page 171: ...Blank Page ...