<5. DENSITY / PARAMETER SETTING>
5-14
IM 11T03E01-51E
2. Alarm-point Setting Mode
[
*ALARM
]
Set the upper /lower limits for the measured values of the parameters noted below, for the
purpose of raising an alarm.
(a) Select a parameter for which alarms are raised.
Physical density [*DENS]
Compensated density [*C_DNS]
Specific gravity
[*SP_GR]
Calorific value
[*CALRY]
Molecular weight [*MOL]
Concentration [*CONCT]
Press the [YES] key to enter the desired mode when it is indicated.
(b) Set the upper/lower limits.
Physical density [*L_DNS], [*H_DNS]
Configure range:
00.0000 to 60.0000 (kg/m
3
),
0.00000
to
4.00000
(lb/ft
3
)
Compensated density [*L_CP.D], [*H_CP.D]
Configure range:
0.0000 to 6.0000 (kg/m
3
),
0.00000
to
0.40000
(lb/ft
3
)
Specific gravity
[*L_SPC], [*H_SPC]
Configure range:
0.0000 to 5.0000
Calorific Value
[*L_CAL], [*H_CAL]
Configure range:
000.000 to 133.000 (MJ/m
3
),
0.0000
to
3.5000
(kBTU/ft
3
)
Molecular weight [*L_MOL], [*H_MOL]
Configure range:
000.00 to 140.00
Concentration [*L_CON], [*H_CON]
Configure range:
000.0 to 100.0