
<6. CALORIE / PARAMETER SETTING>
6-19
IM 11T03E01-51E
CODE 05: Setting of Contact Outputs Status
[*CNTCT
]
This setting defines the status of the contact outputs.
Configurable range: 00 to 15
[ENT]
05
*CODE
*SERVC
XX
*CNTCT
[ENT]
[YES]
F0611.ai
*
1
*
2
*
3
*1:
See "Figure 6.4 Process at the Setting Level" on page 6-12.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
Figure 6.11
Setting of Contact Outputs Status
Each contact output takes either of the following two status (in table); depending on the value you
set.
Value
ZERO contact
SPAN contact MAINT contact
ALM contact
0
NO
NO
NO
NO
1
NO
NO
NO
NC
2
NO
NO
NC
NO
3
NO
NO
NC
NC
4
NO
NC
NO
NO
5
NO
NC
NO
NC
6
NO
NC
NC
NO
7
NO
NC
NC
NC
8
NC
NO
NO
NO
9
NC
NO
NO
NC
10
NC
NO
NC
NO
11
NC
NO
NC
NC
12
NC
NC
NO
NO
13
NC
NC
NO
NC
14
NC
NC
NC
NO
15
NC
NC
NC
NC
NO: means “normally open”
NC: means “normally closed”
Summary of Contents for GD402G /M1
Page 21: ...Blank Page ...
Page 37: ...Blank Page ...
Page 81: ...Blank Page ...
Page 115: ...Blank Page ...
Page 159: ...Blank Page ...
Page 167: ......
Page 169: ......
Page 171: ...Blank Page ...