<5. DENSITY / PARAMETER SETTING>
5-21
IM 11T03E01-51E
CODE 11: Setting for Compensated Density Measurement
This setting defines the reference temperature and pressure required to obtain the compensated
density.
(a) Setting of reference temperature [*C.D.TMP
]
Configurable range:
-20.0 to 80.0 (for °C)
-4.0 to 176.0 (for °F )
(b) Setting of reference pressure [*C.D.PRS
]
Configurable range:
0.10 to 999.99 (for kPa),
0.0001 to 9.9999 (for MPa),
0.015 to 99.999 (for psi )
11
*CODE
*SERVC
XX.X
*C.D.TMP
XXX.XX
*C.D.PRS
[YES]
F0513.ai
*1:
See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
*4: The position of the decimal point changes depending on the setting of CODE 23.
If the code is set to "0" (°C), the reading is XX.X;
if the code is set to "1" (°F), the reading is XXX.X.
[ENT]
[ENT]
[ENT]
*5:
The position of the decimal point changes depending on the setting of CODE 20.
If the code is set to "0" (kPa), the reading is XXX.XX;
if the code is set to "1" (MPa), the reading is X.XXXX;
if the code is set to "2" (psi), the reading is XX.XXX.
*
3
*
5
*
3
*
4
*
2
*
1
Figure 5.12
Setting for Compensated Density Measurement