A2-5
IM 34M06H62-02E
3rd Edition : Jul.16, 2015-00
Table A2.4 Instrument Range and Accuracy (for high resolution operation with SW1-1 set to ON) 3/4
Input
Categor
y
Input
Type
*1
Instrument Range
*2
Input Type Selector Switch
*3
Software
Setting
Accuracy
*4
Resolution
*2
SW1-3 SW1-4 SW5
Software setting (factory setting)
OFF
OFF
0
Instrument ranges may be specified by software
using one of the following codes.
Thermoco
uple
K
*5
-328.0 to 2498.0
F
OFF OFF
1 1
($01)
1.0
F
*5
0.2
F
*5
-328.0 to 1832.0
F
2 2
($02)
1.0
F
*5
0.2
F
*5
-328.0 to 932.0
F
3 3
($03)
1.0
F
*6
0.2
F
*6
J
-328.0 to 2192.0
F
4 4
($04)
1.0
F
*7
0.2
F
*7
-328.0 to 932.0
F
5 5
($05)
1.0
F
*8
0.2
F
T
-454.0 to 752.0
F
6 6
($06)
1.0
F
*9
0.2
F
*9
B
*10
32 to 2912
F
7 7
($07)
2
F
*10
1
F
*10
S
*11
32 to 2912
F
8 8
($08)
2
F
*11
1
F
R
*11
32 to 2912
F
9 9
($09)
2
F
*11
1
F
N
-328.0 to 2372.0
F
A 10
($0A)
1.2
F
*12
0.2
F
*12
E
-454.0 to 1832.0
F
B 11
($0B)
1.0
F
*13
0.2
F
*13
L
-328.0 to 1652.0
F
C 12
($0C)
1.2
F 0.2
F
U
-328.0 to 752.0
F
D 13
($0D)
1.2
F 0.2
F
W
*14
32 to 2912
F
E 14
($0E)
2
F 1
F
Platinel 2
32.0 to 2534.0
F
F 15
($0F)
1.2
F 0.2
F
RT
D
JPt100
-328.0 to 932.0
F
OFF ON
0 16
($10)
0.8
F 0.2
F
-328.0 to 392.0
F
1 17
($11)
0.8
F 0.2
F
32.0 to 572.0
F
2 18
($12)
0.6
F 0.2
F
32.0 to 302.0
F
3 19
($13)
0.4
F 0.2
F
Pt100
-328.0 to 1562.0
F
4 20
($14)
0.8
F 0.2
F
-328.0 to 932.0
F
5 21
($15)
0.8
F 0.2
F
-328.0 to 392.0
F
6 22
($16)
0.8
F 0.2
F
32.0 to 572.0
F
7 23
($17)
0.6
F 0.2
F
32.0 to 302.0
F
8 24
($18)
0.4
F 0.2
F
DC vo
ltage
DC mV
input
*15
0 to 10.00 mV DC
*16
ON
9 25
($19)
0.1% of instrument range
1 digit
*15
0 to 100.0 mV DC
A 26
($1A)
DC V
input
*15
0.000 to 1.000 V DC
B 27
($1B)
0.000 to 5.000 V DC
D 29
($1D)
1.000 to 5.000 V DC
E 30
($1E)
0.00 to 10.00 V DC
F 31
($1F)
*1:
Applicable standard is JIS/IEC/DIN (ITS-90) for thermocouples and RTD.
*2:
For thermocouples K, B, S, R, and W, input ranges may be set wider than their instrument range (see the notes below). However, if the
input range width exceeds 2880
F, the resolution becomes twice the indicated value. Furthermore, the actual range for an acceptable
input is the input range
5%.
*3:
When you turn on the power after changing the hardware switch settings, data stored in the EEPROM is initialized to follow the switch
settings.
*4:
This accuracy applies if the ambient temperature is 77
F
9
F and the input value is within the instrument range. If the input type is
thermocouple and reference junction compensation is used, you should also take into consideration the accuracy of the reference
junction compensation.
*5:
For K-type thermocouples, the input range may be set from -454.0 to 2498.0
F beyond its instrument range. The accuracy and resolution
depend on measured temperatures as follows:
-454.0 to -328.0
F: Neither accuracy or resolution is guaranteed.
-328.0 to 32.0
F:
2.0
F accuracy, 0.4
F resolution
*6:
For K-type thermocouples, the accuracy and resolution depend on measured temperatures as follows:
-328.0 to -292.0
F:
2.0
F accuracy, 0.4
F resolution
-292.0 to -148.0
F:
1.2
F accuracy, 0.2
F resolution
*7:
For J-type thermocouples, the accuracy and resolution depend on measured temperatures as follows:
-328.0 to -148.0
F:
2.0
F accuracy, 0.4
F resolution
*8:
For J-type thermocouples, the accuracy and resolution depend on measured temperatures as follows:
-328.0 to -238.0
F:
1.2
F accuracy, 0.2
F resolution
*9:
For T-type thermocouples, the accuracy and resolution depend on measured temperatures as follows:
-454.0 to -328.0
F:
6.5
F accuracy, 1.0
F resolution
-328.0 to -148.0
F:
2.0
F accuracy, 0.2
F resolution
*10: For B-type thermocouples, the input range may be set from 32 to 3272
F beyond its instrument range. The accuracy and resolution
depend on measured temperatures as follows:
32 to 572
F: Neither accuracy nor resolution is guaranteed.
572 to 1652
F:
5
F accuracy, 1
F resolution
*11: For S-type and R-type thermocouples, the input range may be set from 32 to 3092
F beyond its instrument range. The accuracy and
resolution depend on measured temperatures as follows:
32 to 392
F:
3
F accuracy, 1
F resolution
*12: For N-type thermocouples, the accuracy and resolution depend on measured temperatures as follows:
-328.0 to 32.0
F:
2.5
F accuracy, 0.6
F resolution
*13: For E-type thermocouples, the accuracy and resolution depend on measured temperatures as follows:
-454.0 to -328.0
F:
12.0
F accuracy, 4.0
F resolution
-328.0 to -148.0
F:
2.0
F accuracy, 0.4
F resolution
*14: For W-type thermocouples, the input range may be set from 32 to 4172
F beyond its instrument range.
*15: Resolution is determined by the upper and lower limits for the input range, as well as the upper and lower scaling limits. It is represented
by one digit.
*16: "
" means that the value is ignored.
Summary of Contents for F3CU04-0S
Page 2: ...Blank Page...
Page 18: ...Blank Page...
Page 32: ...Blank Page...
Page 34: ...Blank Page...
Page 50: ...Blank Page...
Page 90: ...Blank Page...
Page 118: ...Blank Page...
Page 130: ...Blank Page...
Page 204: ...Blank Page...
Page 222: ...Blank Page...
Page 224: ...Blank Page...
Page 228: ...Blank Page...
Page 230: ...Blank Page...
Page 232: ...Blank Page...
Page 234: ...Blank Page...
Page 240: ...Blank Page...
Page 242: ...Blank Page...
Page 254: ...Blank Page...
Page 258: ...Blank Page...
Page 260: ...Blank Page...