3-13
IM 710105-03E
Basic Operations
1
2
3
4
5
App
Index
Settings after the Execution of Auto Setup
CH1 to CH4 settings
Position
0 div
Coupling
DC
BW
FULL
Offset
0 V
Invert
OFF
Acquisition settings
Record Length
The same as the value used before you executed auto setup
However, if the record length is such that the DLM2000 cannot
acquire waveforms in Single mode, the record length is set to
the maximum record length at which the DLM2000 can acquire
waveforms repeatedly.
Mode
Normal
Hi Resolution
OFF
Interleave
OFF
Sampling Mode
Interpolation
Trigger settings
Trigger type
Edge
Mode
Auto
HoldOff
20 ns
Delay
0 s
Position
50%
Slope
Rising
Coupling
DC
HF Rejection
OFF
Noise Rejection
OFF
Window
OFF
Settings that depend on the input signal
CH On/Off
On if the DLM2000 detects a voltage of ±20 mV (1:1) or higher and
Off otherwise
V/div
Selects the range with the highest sensitivity that does not exceed
±3.5 divisions
Trigger Level
Center
Trigger Source
The channel with the lowest frequency among the signals whose
amplitude (Max – Min) is at least 1 division
Time/div
The fastest sweep range that allows at least two periods of the
fastest signal among the signals whose amplitude is at least 1
division to be observed. The sweep range must be at least 5 ms/
division.
The values of settings not listed here do not change.
3.5 Performing Auto Setup