Connecting the Power Cord
1.
Check that both the main power switch of the DLM2000 is off.
2.
Connect the power cord plug to the power connector on the rear panel.
3.
Connect the other end of the cord to an outlet that meets the conditions below.
Use the three-prong power outlet equipped with a protective earth terminal.
Rated supply voltage*
100 to 240 VAC
Permitted supply voltage range
90 to 264 VAC
Rated supply voltage frequency
50/60 Hz
Permitted supply voltage frequency range 48 to 63 Hz
Maximum power consumption
Max. approx. 170 VA
* The DLM2000 can use a 100-V or a 200-V system for the power supply. The maximum
rated voltage differs according to the type of power cord. Check that the voltage
supplied to the DLM2000 is less than or equal to the maximum rated voltage of the
provided power cord (see page iii) before using it.
Turning the Instrument On
Before Turning On the Power, Check That:
• The instrument is installed properly:
“2.2 Installing the Instrument”
• The power cord is connected properly: Previous page
Turning On the Main Power Switch
1.
Flip the power switch on the rear panel to the ON ( | ) position.
The front-panel power switch illuminates in red-orange.
Turning On the Power Switch
2.
Press the power switch on the front panel.
The power switch color will change from red-orange to green.
Note
If you turn off the main power switch with the front-panel power switch turned on (with the
power switch illuminated in green), you can start the DLM2000 the next time by simply turning
on the main power switch. However, if you turn off the main power switch while the front-panel
power switch is turned on, the setup data immediately before the power switch is turned off will
not be stored properly.
2.3 Connecting the Power