3-14
IM 710105-03E
3.6 Resetting the DLM2000 to Its Factory Default
Settings
Procedure
Resetting the DLM2000 to Its Factory Default Settings
1.
Press
DEFAULT
key.
The DLM2000 is reset to its factory default settings.
An Undo item appears.
Undoing the Reset Operation
2.
Press the
Undo
soft key to revert to the previous settings.
Explanation
You can reset the DLM2000 settings to their factory default values. This feature is useful
when you want to cancel all of the settings that you have entered or when you want to
redo measurement from scratch.
Settings That Cannot Be Reset to Their Factory Default Values
• Date and time settings
• Communication settings
• Language setting (English or Japanese)
• Measured value font size setting
Undoing the Reset Operation
If you reset the settings by mistake, you can press the Undo soft key to revert to the
previous settings. However, you cannot undo the reset operation if you switch to a
different setup menu or clear the Undo item by pressing the ESC key.
To Reset All Settings to Their Factory Default Settings
While holding down the RESET key ( ), turn the power switch on. All settings are reset
to their factory default values except the date and time settings (the display on/off setting