5.11 General Specifications
Item
Specifications
Standard operating
conditions
Ambient temperature:
23 ± 5°C
Ambient humidity:
55 ± 10%RH
Supply voltage and
frequency errors:
Within 1% of rating
Warm-up time
At least 30 minutes
Storage environment
Temperature:
–20 to 60°C
Humidity:
20 to 80%RH (no condensation)
Altitude:
3000 m or less
Operating environment
Temperature:
5 to 40°C
Humidity:
20 to 80%RH (when the printer is not used; no condensation)
35 to 80%RH (when the printer is used; no condensation)
Altitude:
2000 m or less
Recommended calibration
period
One year
Rated supply voltage
100 to 240 VAC
Permitted supply voltage
range
90 to 264 VAC
Rated supply frequency
50/60 Hz
Permitted supply voltage
frequency range
48 to 63 Hz
Power fuse
Built in (not replaceable)
Maximum power
consumption
170 VA (when the printer is used)
Withstand voltage (between
the power supply and case)
1.5 kVAC for 1 minute
Insulation resistance
(between the power supply
and case)
500 VDC, 10 M
Ω
or more
External dimensions
226 mm (W) × 293 mm (H) × 193 mm (D) when the printer cover is closed; excluding protrusions
Weight
Approx. 4.2 kg (excluding options)
Instrument cooling method Forced air cooling; inlet on the left side, outlet on the rear
Installation position
Horizontal
Battery backup
The clock is backed up with the internal lithium battery.
Battery life: Approx. five years (at an ambient temperature of 25°C)