5-4
IM 710105-03E
Item
Specifications
Selectable trigger delay
range
-(Time length of the post-trigger section) to 10 s
Selectable hold-off time
range
20 ns to 10 s
Trigger type (A trigger)
Edge:
Triggers on the edge of a single trigger source
The source can be set to a signal from CH1 to CH4,
1
from logic bits 0 to 7,
2
EXT, or LINE.
Edge OR:
Triggers when any of the edge trigger conditions of multiple trigger sources is
met
Sources can be set to channels from CH1 to CH4.
1
Edge Qualified:
Triggers on the edge of a single trigger source while qualifications are met
The source can be set to a signal from CH1 to CH4,
1
from logic bits 0 to 7,
2
or
EXT. Qualifications can be specified for CH1 to CH4
1
and for logic bits 0 to 7.
2
State:
Triggers when the state condition changes from met to not met or from not met
to met
State conditions can be applied to all sources using AND or OR.
A clock channel can be specified that is used to determine the met or not met
state condition. No clock can also be selected.
The source and state clock can be set to a signal from CH1 to CH4
1
or from
logic bits 0 to 7.
2
Pulse Width:
Triggers on the width of a single trigger source
The source can be set to a signal from CH1 to CH4,
1
from logic bits 0 to 7,
2
or
EXT.
more than:
Triggers when the time length during which the condition is
met is longer than Time1, and the condition changes to not
met
Time1: 4 ns to 10 s in 2-ns steps
less than:
Triggers when the time length during which the condition is
met is shorter than Time1, and the condition changes to not
met
Time1: 6 ns to 10 s in 2-ns steps
between:
Triggers when the time length during which the condition
is met is longer than Time1 but shorter than Time2 and the
condition changes to not met
Time1: 4 ns to (10 s – 4 ns) in 2-ns steps
Time2: 8 ns to 10 s in 2-ns steps
Minimum spacing between Time1 and Time2: 4 ns
out of range:
Triggers when the time length during which the condition
is met is shorter than Time1 or longer than Time2 and the
condition changes to not met
Time1: 6 ns to (10 s – 4 ns) in 2-ns steps
Time2: 8 ns to 10 s in 2-ns steps
Minimum spacing between Time1 and Time2: 4 ns (2 ns only
when Time1 = 6 ns and Time2 = 8 ns)
time out:
Triggers when the time length during which the condition is
met exceeds Time1.
Time1: 4 ns to 10 s in 2-ns steps
Time accuracy:
4
±(0.5% of s 2 ns)
Minimum time
detection width:
2 ns (typical value
5
)
State Width:
Triggers on the time-duration for which the state condition is met or not met
For a description of the state conditions, see "State Pattern."
The source and state clock can be set to a signal from CH1 to CH4
1
or from
logic bits 0 to 7.
2
For details on the time settings, see "Pulse Width."
FlexRay:
6
Triggers on a FlexRay bus signal
The source can be set to a signal from CH1 to CH4.
Modes:
Frame Start, Error, ID/Data, ID OR
Bit Rate:
2.5 M, 5 M, 10 Mbps
Bus channels: A, B
CAN:
6
Triggers on a CAN (Controller Area Network) bus signal
The source can be set to a signal from CH1 to CH4.
Modes:
SOF, Error, ID/Data, ID OR
Bit Rate:
33.3 k, 83.3 k, 125 k, 250 k, 500 k, 1 Mbps, User Define
For User Define, you can set a value from 10 k to 1 Mbps in 0.1-kbps steps.
5.2 Triggering Section