3-2
SM 701730-01E
3.2
Test Environment
Operating Conditions
• Ambient temperature:
23
±
2˚C
• Humidity:
55
±
10% RH
• Voltage of power supply:
Specified voltage
±
1%
• Frequency of power supply:
Specified frequency
±
1%
Warm Up Time
• More than thrity minutes after tuning ON the instrument.
• Confirm that self calibration is correctly executed after the thirty-minute warm up.
(Be sure to pay attention to the warm up time of all equipment that will be used in the
test.)