2-3
SM 701730-01E
Performance Testing
2
Instruments Used
Model Name
Recommended Device
AC low resistance tester
Kikusui TOS6100
Insulation tester
YOKOGAWA 2407
Withstanding voltage tester
Kikusui TOS-8750
Calibrator
FLUKE (WAVETEK) 9500
Programmable head
FLUKE (WAVETEK) 9520, 5 pc.
BNC cable
50
Ω
terminator
Accessory probe
700988
FET probe
YOKOGAWA 700939
Printer paper
B9850NX
3.5" floppy disk
Formatted 2 HD 1.4 M disk
Flash ATA card
JT**MA3-BD: Fujisoku
PC
Equivalent instruments may be used
Before Testing
• Unless specified within, use the default settings on the main unit during testing.
• The DL1740E and DL1740EL accept 4-channel input, so testing should be performed
on channels 1 through 4. However, the DL1720E has 2-channel input, so testing
should be performed on channels 1 and 2.
• Before testing, restart the unit while holding down the Reset key to initialize the
instrument.
• During testing, also hold down the Reset key when restarting.
2.2.1 Grounding Test
Product specification: 0.1
Ω
or less
Confirm that a current of 30 A flows between the FG inlet of the AC power supply input
and channel 4 of the BNC ground (or between FG and EXT. for the DL1720E), and that
the resistance value is 0.1
Ω
or less.
V-
I-
V+
I+
FG
CH4
AC Low Resistance Tester
DL1720E/DL1740E/
DL1740EL
2.2 Tests for the DL1720E/DL1740E/DL1740EL