2-4
SM 701730-01E
2.2.2 Insulation Resistance Test
Specifications:
For DC 500 V, 10 M
Ω
or higher
Testing Procedure
Measure the insulation resistance between the AC power supply input signals below
when the power switch is turned ON.
DL1720E/DL1740E/
DL1740EL
Power Cable
L
N
FG
Insulation Tester
G
L
N
When the Ethernet (/C10) option is installed, also measure the insulation resistance
across the following signals. (Note: the ETHERNET input terminal should be shorted
across all terminals.)
Insulation Tester
FG
Ethernet Cable
(Terminal is shorted)
DL1720E/DL1740E/
DL1740EL
2.2.3 Withstand Voltage
Testing Procedure
Measure the withstand voltage between the AC power supply input signals below when
the power switch is turned ON.
Power Cable
L
N
FG
Withstanding
Voltage Tester
G
L
N
DL1720E/DL1740E/
DL1740EL
Items to Be Checked
That the product specification is met.
Specification:
AC 1.5 kV for 1 minute
2.2.4 Selftest
The selftest is performed in the STOP condition.
Test the following items.
Testing Procedure
1.
Perform the following test items while referring to “Performing a Self-Test” in the
DL1720E/DL1740E/DL1740EL user’s manual (IM701730-01E).
• Memory Test
• Keyboard Test
• Printer Test (only when the /B5 option is installed)
• Floppy Disk Test (only when the -J1 option is installed)
• PC Card IF Test (only when the -J3 option is installed)
2.
Confirm that all tests were completed normally and that no errors occurred.
3.
Perform the self-calibration, referring to “Performing Calibration” in the
DL1720E/DL1740E/DL1740EL user’s manual (IM 701730-01E).
4.
Confirm that “CAL” in the upper left part of the screen disappears, and that
calibration completed without errors.