3-3
SM 701730-01E
Adjustments
3
3.3
Equipment Required
Table 3.1 Equipment Required
Equipment
Critical Specification
Recommended
Calibrator
DC
FLUKE (WAVETEK) 9500
Programmable Head
Output Level:
1 V
FLUKE (WAVETEK) 9520
Accuracy:
< 0.02%
Square Wave
Frequency:
10 kHz
Output Level:
> 60 V
P-P
Note
The values shown in the specification column are those indicated by this service manual only.
These values do not indicate the actual performances of the recommended equipment and
tools. Therefore, non-designated equipment and tools which satisfy the specifications are
permitted.