17. EXCITATION OUTPUT & POWER SUPPLY
Three isolated transducer excitation output levels are available from the power supply
board. These are selectable via jumpers b, c, d, e, f in the upper right of the board, as
illustrated. In addition, the board provides three jumper positions for special features. The
same jumper locations apply to the universal power supply (85-264 Vac or 90-300 Vdc) and
to the low voltage power supply (12-32 Vac or 10-48 Vdc).
1. Letters indicate jumper position. Jumpers are installed on pins adjacent to letters.
Excitation output
Jumper locations
5 Vdc ±5%, 100 mA max
10 Vdc ±5%, 120 mA max
24 Vdc ±5%, 50 mA max
b, d, e
b, d, f
c
2. SELECTION OF OTHER JUMPERS
Jumper a
- Front panel menu lockout, locked when installed. See Section 9.
Jumper g
- Pr5V power output at P1-4 when installed.
Jumper h
- Connects "Digital Input B" to P1-4 when installed.