YMC Europe GmbH
Cleaning and maintencance
YMC Pilot
39 / 49
PLUS
low pressure chromatography column
10. Adjust the column disc over the rods and lower it onto the glass body.
11. Check that the flat gasket is in the groove in the column disc.
12. Check that the glass body is concentric with the column disc.
13.
ATTENTION
Damage to the glass body is possible due to faulty
assembly. Note the maximum torque during the assembly of the glass
body.
Insert the cylinder screws in a diagonal pattern and tighten them by gradually
increasing the torque to screws on the opposite sides. Adjust to a
maximum
torque of 5 Nm for columns with 70 and 100 mm ID respectively 10 Nm for
columns with 140 and 200 mm ID
.