YMC Europe GmbH
Cleaning and maintencance
YMC Pilot
40 / 49
PLUS
low pressure chromatography column
7.4.3
Frit replacement column body
ATTENTION
Faulty assembly
Damage to the glass body is possible due to faulty assembly.
► Take care during assembly of the glass body that the maximum torque of 3 Nm is
not exceeded.
► Assemble the cylinder screws in a diagonal pattern and always tighten on the
opposite sites.
► Take care during assembly of the glass body that affected parts are correctly
aligned.
Required devices:
• Cylinder key/torque wrench
The frit in the column body is replaced as follows:
1. Remove the adjustable end piece Chapter 6.1, S. 18.
2. Dismantle the glass body Chapter 7.4.2, S. 37.
3. Remove the union nut of the connector.
4. Press the fixed end piece with the connector out of the closing disc (CB).