YMC Europe GmbH
Operation
YMC Pilot
21 / 49
PLUS
low pressure chromatography column
15. Turn the closing disc of the adjustable end piece half turn in the DOWN
direction.
The adjustable end piece is ready for insertion into the column body.
16. Put the piston of the adjustable end piece in a smooth angle opposite the drain
function on the column body.
17. Bring the piston of the adjustable end piece slowly to the vertical position.
Excessive slurry solvent will flow into a reservoir through the drainage tubing.