Specifications and overview
YMC Pilot
9 / 49
PLUS
low pressure chromatography column
3
Specifications and overview
3.1
Fields of application
The YMC Pilot
PLUS
pilot column is especially designed for pilot scale use. The column
is compatible with common BioLC separation modes (IEX, SEC, Affinity, IMAC, HIC,
MMC) and aqueous conditions. Due to the metal-free and the minimal dead volumes,
the column is perfectly suitable for the separation of biomolecules (e.g. proteins,
oligonucleotides or nucleic acids).
3.2
Specifications
PED* fluid group:
Group 1 referred to EU Guideline 2014/68/EU
PED* safety requirement:
Article 4.3 of EU Guideline 2014/68/EU
Column body:
Borosilicate glass 3.3
Calibrated precision glass (Duran KPG®)
Frits:
PE
Seals:
75 Fluoroprene XP41 blue
Connection:
FEP tubing ¼" OD
Wetted parts:
Borosilicate glass 3.3, FEP, PE, PTFE, PEEK, 75
Fluoroprene XP41 blue
Approved temperature range
4
– 40°C
*Pressure Equipment Directive
Length Glass Body
[mm]
500
850
Inner Diameter
[mm]
70
100
140
200
70
100
140
200
Bed Length
[mm]
100 - 450
1
250 - 800
1
Volume [mL]
0.4
–
1.7
0.8
–
3.5
1.5
–
6.9
3.1
–
14.1
1.0
–
3.0
2.0
–
6.2
3.8
–
12.3
7.8
–
25.1
Pressure Limit [bar]
16
10
7
5
16
10
7
5
Total Weight Empty
Column [kg]
19
20
37
38
23
24
45
47
Footprint [mm]
620
750
620
750
Maximum Height [mm]
1,700
1,750
2,250
2,300
1
Depending on packing material and packing method