10. Specifications 

 

Specification of the main unit 

System 

Sealed circulation type 

Circulating gas 

N

2

 gas 

Circulating amount 

0.12

0.65m

3

/min 

Solvent recovery system 

Capacitor(cooling trap

+freezer 

Fine dust collecting system 

Cartridge filter 

Circulating blower 

Roots blower 

Refrigerator 

Air cooling

 

compressor      R404A    800g±10g 

Solvent collecting container 

2L flask 

Indicators 

Trap temperature, oxygen concentration 

Meters 

Filter differential pressure meter, flow meters

for introducing N

2

, for 

the oxygen sensor

 

Filter 

For protecting the O

2

 sensor 

Pump 

For circulation for measuring oxygen concentration 

Compressor 

For pushing the spray nozzle 

Safety device 

Oxygen concentration meter, flammable gas alarm 

Power supply 

AC200V

AC240V

 

Single phase

 

5A 

External dimensions

(㎜)※

W×D×H

) 

700×950×1500 

Weight 

Approx.130kg 

Accessories 

Specimen tube    Silicone: I.D. 2

×O.D. 4

×1m 

               

Viton: I.D. 2

×O.D.4

×1m 

 

Bellows:    40KF L=1300 

 

Bellows:    40KF L=1000 

 

Clamp    40KF 

 

O-ring    40KF 

 

Duct hose    φ25×2m 

 

Hose band    TM-144SUS №5713 

 

Tetrone braided hose    φ6×φ11×5m 

 

Tetrone braided hose    φ6×φ11×1.5m 

 

Fluoro-rubber tube      E-PD-6 φ6×φ9 1.5m 

 

Hose band    TM145-SUS No.5956 

 

Connector    RNJC-PFPM-20-12-A-1 

 

2L collecting flask    GAS41-30330 

 

Flask clamp    KS35 

 

Tube    GAS410_01_02_05 

 

Screw      M4×10 SUS 

 

Warranty card 

 

Operating manual    GAS410 

 

Filter

 

 

Filter element neal 

1    External dimensions do not include protruding parts. 

Summary of Contents for GAS410

Page 1: ...Ltd To use this unit properly read this Instruction Manual thoroughly before using this unit Keep this instruction manual around this unit for referring at anytime WARNING Carefully read and thorough...

Page 2: ...About applicable organic solvents 24 Corrosion resistance table 26 Piping system diagram 27 List of materials 28 Precautions during operation 29 6 Maintenance Method 31 Daily Inspection and Maintenanc...

Page 3: ...equire hospitalization or outpatient visits for a long time Note 2 Minor injury means a wound or an electrical shock that does not require hospitalization or outpatient visits for a long time Note 3 P...

Page 4: ...Hazard of explosion Caution General cautions Electrical shock Burning Caution for no liquid heating Caution for water leak For water only Poisonous material Prohibitions General bans Fire ban Do not d...

Page 5: ...rn off the power switch of the main unit immediately and turn off the original power source and finally contact to either the dealer you purchased this unit or our sales office Leaving the failure may...

Page 6: ...is flown from the sample tube disconnected from the unit and this failure may cause the indeterminism accident Any people other than the qualified personnel shall never attempt to operate the unit Ta...

Page 7: ...nt of the user 2 Pay attention to the color of each core wire when connecting the power cord Be sure to check that the breaker on the power source equipment side is turned OFF when connecting power co...

Page 8: ...may result Refer to page 47 13 List of Dangerous Substances 5 Do not use explosive or flammable substances Take extreme care for use of an explosive or a flammable solvent Such a solvent may cause an...

Page 9: ...s uses the same power line of this unit If the phenomena might be found change the power line of this unit to the other power line For connecting of the device to the power source ask the dealer that...

Page 10: ...eter N2 flow meter Circulation path filter Gas selector switch N2 bombe connection port Duct connection outlet Cooling fan ELB with over current protector Duct connection port Collecting flask door Fo...

Page 11: ...Filter for oxygen sensor Pump for oxygen sensor For condensing solvent in condenser path Temperature sensor Ball valve Clamp 2L collecting flask Freezer Compressor Filter case Flow meter for introduci...

Page 12: ...pressure between the inlet and outlet of the filter to check the filter status Air amount meter monitor installation port Air amount meter and monitor are optional This port is used for installing opt...

Page 13: ...g process will discharge organic solvents hot air and fine particles at the exhaust port out of the pipe path Connect the supplied discharge duct to the discharge port and discharge to external using...

Page 14: ...ol an option for exclusive use of ADL311 in an operation position lowering by all means and can put height together when I am connected to GAS410 The footstool an option for exclusive use of ADL311 be...

Page 15: ...into the connecting tube to the duct connecting inlet port of GAS410 Put an O ring and securely fix using a clamp for installation C Connect a Fluororubber tube milky white and fix it using a wire cl...

Page 16: ...upper position of the connecting port and remove a set of filter installation parts from the lower position of the connecting port 2 Connect the supplied connecting tube using M4 10 truss screws Conne...

Page 17: ...n O ring and securely fix using a clamp for installation C Connect a Fluororubber tube milky white and fix it using a wire clamp D Connect a tetrone braided hose and fix it using a wire clamp E Connec...

Page 18: ...y moves to the next one after calibration or press NEXT key Any key operation will be invalid until the timer counts becomes 0s When connected to GAS410 the blower of ADL311S or GB210A will not be use...

Page 19: ...in N2 gas concentration control by this solenoid valve Open Close Set the liquid tube as shown in the left diagram Push the pump rod CCW to open the pump head put the liquid sending tube in it and the...

Page 20: ...open the ball vale above the collecting flask to start recovering solvent 8 Readjust dry air amount spray pressure and liquid sending speed so that the inlet and the outlet temperature will be stable...

Page 21: ...ontinue operation for a while to completely recover solvent When solvent contains water content the cooling trap may be frosted and the amount of solvent recovered may decrease If amount of solvent re...

Page 22: ...GAS410 OFF 17 Remove the container holding band and take out the product collecting container When taking out the container take care the powder attached on the back side of the cyclone cover 18 Hold...

Page 23: ...is in operation When the lamp is on the liquid sending pump is operating in the normal direction When the lamp is on the liquid sending pump is operating in the reverse direction When this lamp is on...

Page 24: ...t Blower dial value 50 100 150 0 4 44 73 97 4 5 44 75 98 5 5 45 76 100 7 45 76 101 8 5 46 78 102 10 47 79 103 The table below shows the relationship between dial values of the blower and average circu...

Page 25: ...breaker and the main power Immediately contact a service technician for inspection If this procedure is not followed fire or electrical shock may result Never perform repair work yourself since it is...

Page 26: ...educing the air amount and amount of liquid to send for solvents with a low boiling point such as methylene chloride Example Environmental temperature 25 Amount of liquid to send 1170mL H methylene ch...

Page 27: ...anol Acetic ether Hexane Methanol Chloroform Acetone Methylene chloride o 144 m 139 3 p 138 5 82 3 80 1 78 4 77 1 67 7 64 6 61 2 56 2 40 o 25 m 47 4 p 13 2 88 5 5 114 3 83 6 95 3 97 4 63 5 94 6 97 7 o...

Page 28: ...meter Bond to glass O ring Solenoid valve seal Liquid sending tube Hoses Packing for installing the oxygen concentration measurement sensor Filter bottom packing Oxygen concentration meter pump valve...

Page 29: ...27 5 Handling precautions Piping system diagram...

Page 30: ...onnecting hose Teflon G Cap Silicone Compressor Cover case PP Connecting hose Teflon H Pressure meter Solenoid valve for N2 control Seal assembly Teflon I Needle valve Seal assembly Teflon Viton Flow...

Page 31: ...f the cyclone without falling into the container for product Therefore pay special attention to the unit with the container for product be mounted 11 When the heater is ON do not expose the cap and ba...

Page 32: ...electricity such as attaching included earth clips at three positions on the clamp at the connection of the cyclone or attaching an antistatic brush to the body of the cyclone 18 Depending on the samp...

Page 33: ...normal one Do not disassemble this unit Caution Use a well drained soft cloth to wipe dirt on this unit Do not use benzene thinner or cleanser for wiping Do not scrub this unit Deformation deteriorati...

Page 34: ...filter element and remove dusts attached inside the filter case or on the bottom by suctioning with an electric cleaner or by wiping out with a cloth During this work take care so that dusts will not...

Page 35: ...three months thereafter Remove the blind plates on the right side plate and at the lower right of the rear side Stop the blower remove the oil drain plug and allow all old oil to discharge Replace th...

Page 36: ...sassemble this unit as possible and recycle the reusable parts considering to the environmental protection The feature components of this unit and materials used are listed below Component Name Materi...

Page 37: ...p temperature becomes 40 or more Imperfect connection or disconnection of the temperature sensor Malfunction of the display monitor temperature controller Malfunction of the freezer or the inverter Po...

Page 38: ...ly voltage Repair or replace Lower ambient temperature or cleaning the plug to ensure a well ventilated Repair or replace Supply voltage too high or too low Motor mechanical failure Ambient temperatur...

Page 39: ...onfirmation and language select for the manual You can select English or Chinese with the language select function for the manual Confirmation and language select for the manual Pressing ESC key will...

Page 40: ...and language select for the manual You can select English or Chinese with the language select function for error display and comments Language selection for error display Language selection for commen...

Page 41: ...efective Wait for five minutes and check whether the freezer operates or not Replace the heater Replace the touch panel or the sequencer Repair the defective points or replace the temperature controll...

Page 42: ...connect the test port of oxygen concentration meter to atmosphere directly See P 40 picture Replace the flowmeter Replace Replace the filter Replace Replace Thermo regulator failure Defective display...

Page 43: ...arbon filter absorption operate as per the following pictures Adjust the gas pipe switching valve to connect to atmosphere External gas wait for several minutes and check if the oxygen concentration d...

Page 44: ...nd other information and send it to our customer service center by Facsimile 03 3231 6523 Then store it securely Warranty period is one full year from the date of purchase Repair service for free is a...

Page 45: ...ntration Compressor For pushing the spray nozzle Safety device Oxygen concentration meter flammable gas alarm Power supply AC200V AC240V Single phase 5A External dimensions 1 W D H 700 950 1500 Weight...

Page 46: ...44 11 Wiring diagram...

Page 47: ...N2 YSJ A040403013 Solenoid valve CKD AB41 02 5 F AC100V for controlling N2 YSJ A040403004 Linear compressor AC0910 YSC B030200001 Differential pressure meter DG87 641 1C 0 1kPa YSC B042300001 Filter...

Page 48: ...Temperature controller TTM214 Q RI RUV for trap display YSC B020101027 Relay HF13F A1002Z1D YSJ A011002005 Transformer AD21 200A2 YSC B010701011 Oxygen sensor module GPR 2500 GPR11 32 RST YSJ B9901002...

Page 49: ...chlorate ammonium perchlorate and other perchlorates Potassium peroxide sodium peroxide barium peroxide and other inorganic peroxides Potassium nitrate sodium nitrate ammonium nitrate and other nitrat...

Page 50: ...s Unexpected accidents or breakdown may result in Note The contents of this document may be changed in future without notice Any books with missing pages or disorderly binding may be replaced Instruct...

Reviews: